General description:
The MultiPhoton Imager-121 is an
ultra-low background gamma imager for low-activity distributed 125I
gamma and 131I sources. Measurements are performed simultaneously in
two modes: standard (faster), and enhanced (slower, but more sensitive).
The instrument can measure the 20 gels which have dual
labels,125I and 131I.
Components: MPD Imager-121 includes a multi-element detector unit with
a positioning table, a dedicated computer with mouse and keyboard, and a 15"
color monitor.
Sample Size: Imaged area up to 25 x 25 cm.
Resolution: Better than 0.3 mm
Sensitivity: 0.15 dpm/pixel (30 zeptomole of 125I) in
standard mode, and 0.006 dpm/pixel (1.2 zeptomole of 125I) in
enhanced mode. Similar sensitivity has been achieved for 131I
Dynamic Range: From 3 attoCi up to at least 5 nCi per pixel: more than 6
orders of magnitude.
Linearity: Better than 5% across the full dynamic range.
Software: Imager 1.0 software running under Windows 3.1 for measurement.
Imgview 1.0 software running under Windows 3.1 and Windows 95 for analysis and
plotting the data. Biotraces 1.0 wavelet based data processing/image cleaning
software, running under Windows 95
Power Requirements: 110 V, 60 Hz, grounded outlet. Power consumption 300 W.
A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40° C.
Humidity: 10 to 95% non-condensing.
Weight: Detector unit: 40 kg, Computer in industrial-type case: 10 kg.
Dimensions: Detector unit: 24" x 30" x 21", Computer unit: 19" x 18" x 7".
Price: $55,000.00, including a one-year warranty and service and two
years of free software upgrade.