srMPD/DB-225: Specifications (July 1999)
General description:
The Spatially Resolving MultiPhoton Detector for dot blot patterns (srMPD/DB-225)
is an ultra-low background gamma counter for low-activity 125I
gamma sources. Measurements are performed simultaneously in two modes: standard
(faster), and enhanced (slower, but more sensitive).
Components: srMPD/DB-225 includes a multi-element detector unit
, a formatted cassette in which the sample is inserted, a dedicated computer
with mouse and keyboard, and a 17" color monitor.
Sample Size: The accepted sample is a 1.5" x 1.5" dot
blot membrane of 225 dots arrayed in 15x15 dots pattern.
Sensitivity: 2 pCi (1 attomole of 125I) in standard mode, and
0.1 pCi
(0.05 attomole of 125I) in enhanced mode *.
Dynamic Range: From 0.1 pCi up to at least 30 nCi: close to 6 orders of
magnitude.
Linearity: Better than 5% across the full dynamic range.
Throughput: 225 samples are measured at the same time.
Software: MPS 1.0 software running under Windows 3.1
Power Requirements: 110 V, 60 Hz, grounded outlet. Power consumption 300 W.
A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40°C. Humidity: 10 to 95%
non-condensing.
Weight: Detector/changer unit: 40 kg. Computer: 10 kg.
Dimensions: Detector/changer unit: 22" x 20" x 21". Computer unit: 19" x
18" x 7".
Price: $ 60,000 including a one-year warranty and service and two years of
free software upgrade.
* Using BioTraces Inc. proprietary SuperTracers the sensitivity of about
20 molecules can be achieved