srMPD/MT-384: Specifications (July 1999)
General description: The Spatially Resolving MultiPhoton Detector for Microtiter wells (srMPD/MT-384) is an ultra-low background gamma counter for low-activity 125I gamma sources. Measurements are performed simultaneously in two modes: standard (faster), and enhanced (slower, but more sensitive).
Components:
srMPD/MT-384 includes a multi-element detector unit with
2-dimensional sample changer for 384 wells microtiter plate, working in
an automatic regime,and includes a dedicated computer with mouse and keyboard,
and a 17" color monitor.
Sample Size: 384 well microtiter plate.
Sensitivity: 2 pCi (1 attomole of 125I) in standard mode, and
0.1 pCi (0.05 attomole of 125I) in enhanced mode * .
Dynamic Range: From 0.1 pCi up to at least 30 nCi: about 6 orders of magnitude.
Linearity: Better than 5% across the full dynamic range.
Throughput: 64 samples are measured at the same time at one position of
2-dimensional sample changer.
Software: MPData 1.0 software running under Windows 3.1
Power Requirements: 110 V, 60 Hz, grounded outlet. Power consumption 300 W.
A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40° C. Humidity: 10 to 95%
non-condensing.
Weight: Detector/changer unit: 40 kg. Computer: 10 kg.
Dimensions: Detector/changer unit: 22" x 20" x 21". Computer unit: 19" x
18" x 7".
Price: $ 50,000 including a one-year warranty and service and two years
of free software upgrade.
* Using BTI Inc. proprietary SuperTracers the sensitivity of about 20 molecules can be achieved.