
General description:
The Spatially
Resolving MultiPhoton Detector for dot blot patterns (srMPD-DB-96)
is an ultra-low background gamma counter for low-activity 125I
gamma sources. Measurements are performed simultaneously in two modes: standard
(faster), and enhanced (slower, but more sensitive).
Components: srMPD/DB-96 includes a multi-element detector unit
with 2- dimensional sample changer
, a dedicated computer with mouse and keyboard, and a 17" color monitor.
Sample Size: srMPD/DB-96 rectangular 110 × 75 mm dot blot membrane of 96
dots arrayed according to the standard 96-well microtitration plate format.
srMPD-DB-96 measures the 96-well microtiter plates.
Sensitivity : 2 pCi (1 attomole of 125I) in standard mode,
and 0.1 pCi (0.05 attomole of 125I) in enhanced mode *.
Dynamic Range: From 0.1 pCi up to at least 30 nCi: close to 6 orders of
magnitude.
Linearity: Better than 5% across the full dynamic range.
Throughput: 16 samples are measured at the same time at one position
of 2-dimensional sample changer
Software: MPS 1.0 software running under Windows 3.1
Power Requirements: 110 V, 60 Hz, grounded outlet.
Power consumption 300 W. A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40°C. Humidity:
10 to 95% non-condensing.
Weight: Detector/changer unit: 40 kg. Computer: 10 kg.
Dimensions: Detector/changer unit: 22" x 20" x 21".
Computer unit: 19" x 18" x 7".
Price :$45,000 including a one-year warranty and two years of free
software upgrade.
* Using BioTraces Inc. proprietary SuperTracers the sensitivity of about
20 molecules can be achieved