General description:
The Spatially
Resolving MultiPhoton Detector for Microtiter wells (srMPD/MT-96) is an
ultra-low background gamma counter for low-activity 125I gamma sources.
Measurements are performed simultaneously in two modes: standard (faster), and
enhanced (slower, but more sensitive).
Components: srMPD/MT-96 includes a multi-element detector unit with
2-dimensional sample changer for 12 strips of 8 samples each, working in an
automatic regime, a dedicated computer with mouse and keyboard, and a 17" color
monitor.
Sample Size: 12 strips with 8 samples each with volume less than 50 ul per
sample in standard 9 mm x 12 mm wells.
Sensitivity: 2 pCi (1 attomole of 125I) in standard mode,
and 0.1 pCi (0.05 attomole of 125I) in enhanced mode * .
Dynamic Range: From 0.1 pCi up to at least 30 nCi: about 6 orders of
magnitude.
Linearity: Better than 5% across the full dynamic range.
Throughput: 16 samples are measured at the same time at one position of
2-dimensional sample changer.
Software: MPData 1.0 software running under Windows 3.1
Power Requirements: 110 V, 60 Hz, grounded outlet. Power consumption
300 W. A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40° C.
Humidity: 10 to 95% non-condensing.
Weight: Detector/changer unit: 40 kg. Computer: 10 kg.
Dimensions: Detector/changer unit: 22" x 20" x 21".
Computer unit: 19" x 18" x 7".
Price: $ 45,000 including a one-year warranty and two years of free
software upgrade.
* Using BTI Inc. proprietary SuperTracers the sensitivity of about
20 molecules can be achieved.