
General description:
The sequential sample MultiPhoton Detector (ssMPD-12)
is an ultra-low background gamma counter for low-activity 125
I gamma sources. Measurements are performed simultaneously in two modes:
standard (faster) and enhanced (slower, but more sensitive).
Components: ssMPD-12 includes a multi-element detector unit with sample
changer for 16 samples working in an automatic regime, a dedicated computer with
mouse and keyboard, and a 15" color monitor.
Sample Size: 16 samples with volume less than 2 ml per sample in standard
12 mm x 75 mm polyethylene tubes.
Sensitivity: 1 pCi (0.5 attomoles of 125I) in standard mode
and 0.1 pCi (50 zeptomoles of 125I) in enhanced mode*.
Dynamic Range: 0.1 pCi to at least 100 nCi: greater than 6 orders of
magnitude.
Linearity: Better than 5% across the full dynamic range.
Software: MPData 1.0 software running under Windows 3.1
Power Requirements: 110 V, 60 Hz, grounded outlet. Power consumption
200 W. A surge protector is recommended.
Environmental Requirements: Temperature: 15 to 40° C.
Humidity: 10 to 95% non-condensing.
Weight: Detector/changer unit: 40 kg. Computer in industrial-type case:
10 kg.
Dimensions: Detector/changer unit: 22" x 11" x 22".
Computer unit: 19" x 18" x 7".
Price:$30,000 including a one-year warranty and two years of free
software upgrade.
* Using BTI Inc. proprietary SuperTracers the sensitivity of about
20 molecules can be achieved